top of page
Integrated Type 1,2,3 Water Purification System

Integrated Type 1,2,3 Water Purification System

Our Dura-Q Premium Series of water purification systems offers an economical Type-I (Ultrapure grade), Type II (Analytical grade), Type-III (Lab grade) water in a single integrated table top/wall mounted sleek unit.

 

Download Brochure 

Highlights:

  • Three stage Integrated pretreatment module of 10µ/5µ/1µ/0.5µ/iron filter, activated carbon, antiscalant,enough to protect and enhance the life of the precious cartridges.
  • High recovery upto 60% of water with polyamide T.F.C.membrane double RO Pass
  • Most economical ultra pure water purification system comparing to other systems in the market
  • Ionic and Organic Rejection/Removal: >99%

 

Feed Water Specifications:

 Potable Tap water with pressure  0.5 kg to 1.5 Kg (1-6 bar or better),Conductivity: upto 2000µS/cm, Chlorine:<3ppm, SDI <12, TOC<2000ppb, CO2: < 50ppm, Silica  <30ppm, Temperature: 4 to 40 Degree C,  Flow rate: up to 75 L/hr, Drain: > 90L/hr, System  include Pre-Filtration Cartridge (1,5 & 10 µ ) with Booster Pump and 01 Bio filter (point-of-use Biofilter of 0.2micron) to remove residual biologically active contaminants

 

Specifications:

 

Water Purification System Model:

Dura-Q Premium

Display 

5.7” TFT /LED/LCD

Display

Product Water 

Type-I/Type-II/Type-III

Type-I Grade Water

Specifications:-

 

Resistivity 

18.2 MOhm.cm

at 25°C

Conductivity

0.05µS/cm

TOC

<5ppb*

Particulates

(>0.1µm&<0.22µm)

<1/ml

Bacteria

(1cfu/100ml)

<0.01cfu/ml

Endotoxins/

Pyrogens (EU/ml)

<0.001

RNase (ng/ml)

<0.001

DNase (pg/µl)

<0.005

Dispense Rate

(Programmable)

> 2L/min

UF Membrane

(NMWL/MWCO)

5000

End Point Filter

0.2,0.1µ

(EFA: 250cm/260cm )

Type-II Grade Water

Specifications:-

 

Resistivity

10-15 MOhm.cm

at 25°C

Conductivity

0.1- 0.06µS/cm

Dispense Rate

(Programmable)

upto 2L/min

TOC

<20ppb***

Particulates

(>0.1µm &<0.2µm)

<1/ml

Bacteria (cfu/10ml)

<0.1cfu/ml

Deionization

Mixed Bed

Resins/EDI****

Type-III Grade Water

Specifications:-

 

Conductivity

<20µS/cm***

Product Flow Rate

(L/hr)

5/10/15/20/25/30

Tank

Integrated/ Inbuilt

Feed/RO/%Rejection

MonitoringYes

Yes

Cell Constant

 

K= 0.01/0.1/1.0

bottom of page